Çalışma Saatleri : 09:00 - 18:00

OctoStat

8-channel Potentiostat/Galvanostat/ZRA

 19″ rack-mountable housing

 Up to 128 channels/rack

 Affordable high-spec electrochemical test system

 Each channel: Dedicated embedded FRA/EIS

Dedicated application

The OctoStat is designed for applications where multiple cells are operated in parallel for long periods of time, such as battery/fuel cell-testing, electrode development, biotechnology and corrosion.

Complete solution

The OctoStat offers a complete package. The hardware includes a built-in dedicated Frequency Response Analyser for each channel, low current measurement possibility down to 100pA full-scale with fA resolution, and an analog measurement channel for secondary signals such as temperature. All measurement and dataprocessing software is included.

Safety

The maximum current or potential of the instrument can be defined by the operator. Thus valuable samples may be protected, and unsafe situations prevented.

Data security & Connection

The OctoStat is equipped with an integrated DataSecure module:

  • Data is stored on board independent of the computer.
  • Connection of multiple PCs is possible.
OctoStat30OctoStat200OctoStat5000OctoBoost16000
System performance
Current compliance±30mA±200mA±5A±16A
Maximum compliance voltage±10V±10V±10V-2 to +9V, or ±5V (user selectable)
Maximum applied voltage±10V±10V±10V-2 to +9V, or ±5V (user selectable)
Potentiostat
Applied potential range±10V, 0.08mV res.±10V, 0.08mV res.±10V, 0.08mV res.
Applied potential accuracy0.2%, or 2mV0.2%, or 2mV0.2%, or 2mV
Current ranges±100pA to ±10mA±100pA to ±100mA±100pA to ±10A±10A and ±100A
Galvanostat
Applied current accuracy0.2%0.2%0.2%
Galvanostatic current ranges±10nA to ±10mA±10nA to ±100mA±10nA to ±10A
Impedance Analyser
Frequency range10µHz to 100kHz (1MHz optional)10µHz to 100kHz (1MHz optional)10µHz to 100kHz (1MHz optional)10µHz to 10kHz
Environment
InterfacingUSB/LAN (Ethernet)USB/LAN (Ethernet)USB/LAN (Ethernet)
Module height1U1U2U3U
Integrated DataSecureData acquisition time: 2ms minimum, Stored no. of data points: 20M each channelData acquisition time: 2ms minimum, Stored no. of data points: 20M each channelData acquisition time: 2ms minimum, Stored no. of data points: 20M each channelData acquisition time: 2ms minimum, Stored no. of data points: 20M each channel
PC requirementsWindows 8/10/11,
with free USB port
Windows 8/10/11,
with free USB port
Windows 8/10/11,
with free USB port
Windows 8/10/11,
with free USB port
  • Automatic current range selection: 10nA to 100mA/10A
  • Automatic variable noise filter and potentiostat/galvanostat stability settings
  • Simultaneous acquisition of current/potential and peripheral analog signal
  • Real time data up to 3,000 pnts/s. Acquisition up to 300,000 pnts/s is stored in instrument memory
  • Minimum interval time: 3.33µs (0.125µs resolution)
  • Result optimalisation by automated tuned filters, gain amplifiers and DC-subtraction

The OctoStat is controlled via Ivium’s own software, IviumSoft. This versatile, yet intuitive software, allows instrument control, data management and analysis, etc. The OctoStat can also be controlled from LabVIEWTM, C++, Delphi, etc. A full suite of software is included as standard with each instrument.

  • Control of separate channels individually
  • All channels simultaneously
  • Synchronous start
  • Data from different channels plotted individually or in 1 single graph