Çalışma Saatleri : 09:00 - 18:00
Category

ZENNIUM XC

HIGHEND MEETS COMPACT

  • DC voltage ranges: ±5 V and ±14 V
  • Current up to ±2 A over 12 current ranges
  • EIS frequency range from 10 µHz to 5 MHz
  • AC amplitude of 0 – 6 V or 0 – 2 A for EIS
  • Switchable floating / grounded mode

The ZENNIUM XC is a compact potentiostat / galvanostat including a frequency response analyser. It offers 32 bit DC resolution and 24 bit AC resolution across 12 different current ranges, which ensure exceptional accurate and precise measurements. The ZENNIUM XC combines compact design with the high-end measurements hardware.

ZENNIUM XZENNIUM PROZENNIUM XC
Electrochemical impedance spectroscopy (EIS)Potentiostatic EIS
Galvanostatic EIS
Psuedo-galvanostatic EIS
EIS at OCP
Series EIS measurements vs parametersVoltage/current/time etc
Single frequency impedance vs parameterVoltage/current/time etc
Multichannel EIS measurements(with PAD4 cards)up to 17 channelsup to 5 channels-
Testsampling online display and logging
Current voltage curveSteady state sampling
Fixed sampling
Dynamic sampling
Tafel scan
Polarization measurementsPotentiostatic polarization
Galvanostatic polarization
OCP measurement
Linear sweep voltammetry (LSV)
Linear sweep amperometry
Chronopotentiometry
Chronoamperometry
Chronocoulometry
Cyclic voltammetry (CV)
Parallel CV measurements(with PAD4 cards)up to 17 channelsup to 5 channels-
Time domain measurements
PolarographyTastpolarography
Differential pulse polarography
Differential pulse voltammetry
Square wave voltammetry
Constant potential pulse voltammetry
Battery cycling(with or without EIS)
PITT / GITT
Automation via SCRIPT
Remote control via Python or Labview
Bi-potentiostatwith XPOT2with XPOT2-
Sequential multicell measurementswith multiplexer extensions-
Recording of additional analog or digital in- & output signalstemperature, voltage channels-
Electrochemical noise measurementswith Noise Probewith Noise Probe-
Hardware extension for photoelectrochemical measurements-
Hardware extension for photovoltaic measurements-
Hardware extensions for high current measurements-
AC-DC-AC testwith COLTwith COLTwith COLT
Layer quality test or Bilayer testwith COLTwith COLTwith COLT
ZENNIUM XZENNIUM PROZENNIUM XC
General
Overall bandwidthDC - 15 MHzDC - 10 MHzDC - 6 MHz
ADC resolution32 bit32 bit32 bit
Harmonic reject> 60 dB @ ½ full scale> 60 dB @ ½ full scale> 60 dB @ ½ full scale
Potentiostat modespotentiostatic, galvanostatic, pseudo-galvanostatic, OCP, ZRA, FRApotentiostatic, galvanostatic, pseudo-galvanostatic, OCP, ZRA, FRApotentiostatic, galvanostatic, pseudo-galvanostatic, OCP, ZRA, FRA
Cell connection2-, 3-, 4-terminal kelvin2-, 3-, 4-terminal kelvin2-, 3-, 4-terminal kelvin
Ground referencegrounded, floatinggrounded, floatinggrounded, floating
Extension slots105
PC interfaceUSB 2.0USB 2.0USB 2.0
Dimensions (H x W x D)160 x 470 x 376 mm3160 x 364 x 376 mm3160 x 255 x 385 mm3
Weight13.2 kg11.4 kg8 kg
Accessoriescell cable set, USB cable, power cord, Testbox, Thales XT software suite, manual, 3 Zahner Analysis licensescell cable set, USB cable, power cord, manual, Thales XT software suite, 3 Zahner Analysis licencescell cable set, USB cable, power cord, manual, Thales XT software suite, 3 Zahner Analysis licenses
Power supply100/115/230VAC, 50/60Hz100/115/230VAC, 50/60Hz100/115/230 VAC, 50/60 Hz
Ambient temperature+10 °C to +30 °C+10 °C to +30 °C+10 °C to +30 °C
Ambient humidity< 60% without derating< 60% without derating< 60% without derating
Storage temperature-20 °C to +85 °C-20 °C to +85 °C​-20 °C to +85 °C
Input
Maximum input voltage±5.5 V
±16 V
5 V range
15 V range
±5.5 V
±16 V
5 V range
15 V range
±5 V
±14 V
5 V range
14 V range
Voltage resolution2.5 nV
7.5 nV
5 V range
15 V range
2.5 nV
7.5 nV
5 V range
15 V range
2.5 nV
7.5 nV
5 V range
14 V range
Voltage accuracy±20 µV ± 2 ppm of reading
±50 µV ± 5 ppm of reading
5 V range
15 V range
±50 µV ± 5 ppm of reading
±100 µV ± 10 ppm of reading
5 V range
15 V range
±100 µV ± 5 ppm of reading
±300 µV ± 10 ppm of reading
5 V range
14 V range
Current resolution±2 aA (32 bit)
Main / HiZ Probe
±2 aA (32 bit)
Main / HiZ Probe
±2 aA (32 bit)
Main / HiZ Probe
Current accuracy±0.05% of reading ± 0.01% of FS
±0.5% of reading ± 0.1% of FS
± 0.5% of reading, ± 125 fA
1 µA ... 100 mA
100 mA ... 4 A
10 nA ... 1 µA
< |1 nA| HiZ Probe
±0.05% of reading ± 0.02% of FS
±0.5% of reading ± 0.2% of FS
± 0.5% of reading, ± 125 fA
1 µA ... 100 mA
100 mA ... 3 A
1 nA ... 1 µA
< |1 nA| HiZ Probe
±0.05% of reading ± 0.04% of FS
±0.5% of reading ± 0.4% of FS
± 0.5% of reading, ± 125 fA
1 µA ... 100 mA
100 mA ... 2 A
10 nA ... 1 µA
< |1 nA| HiZ Probe
Input impedance> 10 TΩ || ±5 pF (typical)
> 1000 TΩ || ±1 pF (typical)
Main
HiZ Probe
> 10 TΩ || ±5 pF (typical)
> 1000 TΩ || ±1 pF (typical)
Main
HiZ Probe
> 10 TΩ || ±5 pF (typical)
> 1000 TΩ || ±1 pF (typical)
Main
HiZ Probe
Input leakage current< |200 fA| typ.
< |2 pA| max.
|10 fA| typ.
HiZ Probe
< |200 fA| typ.
< |2 pA| max.
|10 fA| typ.
HiZ Probe
< |200 fA| typ.
< 2 - 5 pA
< |10 fA| typ.
Main
HiZ-Probe
Impedance range1 mΩ to 10 GΩ / 1%
100 mΩ to 10 MΩ / 0.1%
100 mΩ to 100 GΩ / 2%
10 µΩ to 1 GΩ / 2%
1 mΩ to 10 MΩ / 0.2%
Main
HiZ
Gal
1 mΩ to 10 GΩ / 2%
100 mΩ to 10 MΩ / 0.2%
100 mΩ to 100 GΩ / 2%
10 µΩ to 1 GΩ / 2%
1 mΩ to 10 MΩ / 0.2%
Main
HiZ
Gal
1 mΩ to 10 GΩ / 2%
100 mΩ to 10 MΩ / 0.2%
100 mΩ to 100 GΩ / 2%
10 µΩ to 1 GΩ / 2%
1 mΩ to 10 MΩ / 0.2%
Main
HiZ
Gal
Common mode rejection> 86 dB @ 10 µHz to 100 kHz
> 66 dB @ 100 kHz to 12 MHz
> 86 dB @ 10 µHz to 100 kHz
> 66 dB @ 100 kHz to 8 MHz
> 86 dB @ 10 µHz to 100 kHz
> 66 dB @ 100 kHz to 5 MHz
Input channel phase-tracking accuracy±0.05 ° @ 10 µHz to 100 kHz
±0.125 ° @ 100 kHz to 12 MHz
±0.05 ° @ 10 µHz to 100 kHz
±0.125 ° @ 100 kHz to 8 MHz
±0.05 ° @ 10 µHz to 100 kHz
±0.125 ° @ 100 kHz to 5 MHz
Equivalent effective input noise1 µV rms / 100 fA rms @ 1 mHz to 10 Hz
10 µV rms / 15 fA rms @ 1 mHz to 10 Hz
Main
HiZ
1 µV rms / 100 fA rms @ 1 mHz to 10 Hz
10 µV rms / 15 fA rms @ 1 mHz to 10 Hz
Main
HiZ
1 µV rms / 100 fA rms @ 1 mHz to 10 Hz
Main
Output potentiostatic
Controlled voltage±5 V
±15 V
±5 V
±15 V
±5 V
±14 V
Resolution2.5 nV
7.5 nV
5 V range
15 V range
2.5 nV
7.5 nV
5 V range
15 V range
2.5 nV
7.5 nV
5 V range
14 V range
Accuracy±50 µV ± 5 ppm of reading
±200 µV ± 5 ppm of reading
5 V range
15 V range
±100 µV ± 10 ppm of reading
±500 µV ± 10 ppm of reading
5 V range
15 V range
±200 µV ± 10 ppm of reading
±600 µV ± 10 ppm of reading
5 V range
14 V range
Integral Nonlinearitytyp. 1 ppm, max. 2 ppm
typ. 3 ppm, max. 8 ppm
5 V range
15 V range
typ. 2 ppm, max. 4 ppm
typ. 6 ppm, max. 12 ppm
5 V range
15 V range
typ. 4 ppm, max. 8 ppm
typ. 12 ppm, max. 24 ppm
5 V range
14 V range
Compliance voltage±16 V
±32 V
±14 V
±28 V
Low compliance voltage mode
High compliance voltage mode
AC amplitude0 to 2 V, 24 bit resolution
0 to 6 V, 24 bit resolution
5 V range
15 V range
0 to 2 V, 24 bit resolution
0 to 6 V, 24 bit resolution
5 V range
15 V range
0 to 2 V, 24 bit resolution
0 to 6 V, 24 bit resolution
5 V range
14 V range
Bandwidth15 MHz @ 33 Ω load10 MHz @ 33 Ω load6 MHz @ 33 Ω load
IR compensationauto AC impedance technique
range 0 to 10 MΩ
resolution 0.012%
auto AC impedance technique
range 0 to 10 MΩ
resolution 0.012%
auto AC impedance technique
range 0 to 10 MΩ
resolution 0.012%
Small signal rise time150 ns to 200 µs in 5 steps
automatic selection
150 ns to 200 µs in 5 steps
automatic selection
150 ns to 200 µs in 5 steps
automatic selection
Slew rate15 MV/s15 MV/s15 MV/s
Phase shift10° @ 500 kHz10° @ 500 kHz10° @ 500 kHz
Output galvanostatic
Maximum current±4 A±3 A±2 A
Current range±1.9 nA to ±4 A in 12 steps
Main
±1 nA to ±3 A in 12 steps
Main
±1 nA to ±2 A in 12 steps
Main
Resolution32 bit, ±0.2 ppb of FS32 bit, ±0.2 ppb of FS32 bit, ±0.2 ppb of FS (1ppm @ 1 pA)
Accuracy±0.025% of reading ± 0.01% of FS
±0.1% of reading ± 0.05 % of FS
≥ 1 µA to 100 mA
< 1 µA or > 100 mA
±0.05% of reading ± 0.02% of FS
±0.2% of reading ± 0.1% of FS
≥ 1 µA to 100 mA
< 1 µA or > 100 mA
±0.1% of reading ± 0.04 % of FS
±0.4% of reading ± 0.2% of FS
≥ 1 µA to 100 mA
< 1 µA or > 100 mA
Frequency generator & analyzer
Frequency range10 µHz to 12 MHz10 µHz to 8 MHz10 µHz to 5 MHz
Accuracy< 0.0025%< 0.0025%< 0.0025%
Resolution0.0025%,
10000 steps/decade
0.0025%,
10000 steps/decade
0.0025%,
10000 steps/decade