The Cell-T module allows temperatures to be measured and recorded simultaneously with the primary signal, without the need for a peripheral port. This brings extra functionality to instruments such as the dModules in the Ivium-n-Stat and the Vertex.One/Vertex.C. The module connects between the instrument and the cell cable and connection of the thermocouple is via a K-type connector. The temperature is automatically recorded on the analog input 1 channel.
- Suitable for K-type thermocouples
- The input of the thermocouple module is not isolated from the potentiostat’s ground. If isolation of the thermocouple is required, a thermocouple with isolated tip should be used
- Powered from the cell connector
- Temperature range: 10 to 900C
- Measurement accuracy:
- 10 to 400C is ±2C
- >400C is ±1.6%
- Output signal: analog output voltage of 5mV/C
- Port interfacing: the voltage signal transfers to analog input1 of the potentiostat