The sensor is formed on a corundum ceramic base. On to this surface the working, the reference and the auxiliary electrodes are applied. Working electrode has homogenous surface with roughness less than 1 µm. The working and the auxiliary electrodes are made of variety of materials. At the end of the sensor there is a contacting field which is connected with the active part by the silver conducting paths which are covered by a dielectric protection layer. A bio-chemically active substance can be immobilised on the working electrode of the sensor to create a biosensor.
AC*. W*. R* (*)
AC – amperometric sensor or electrode on corundum ceramic base
AC*– Sensor group reference number: 1-13
CC*. W* (*)
CC – conductometric electrode on corundum ceramic base
CC*– Sensor group reference number: 1-3
W* – Working electrode material
S – Alloy of Gold and Platinum
1 – Pure gold
2 – Pure platinum
3 – Pure silver
4 – Carbon(Graphite)
5 – Manually Microdispensed Carbon(Graphite) with Au+Pt alloy auxiliary electrode
R* – Reference electrode material
S – Silver
1 – Silver/Silver Chloride
2 – Silver covered by AgCl
(*) – Additional Technical specification
H – Heating of the sensor
T – Temperature sensing element