Sensor Usage: Amperometric measurement of β-D-Glucose concentration in liquid samples (from 0 °C to 45 °C)
Glucose Oxidase (GOD) from Aspergillus Niger is immobilized on the active surface of a working electrode of amperometric substrate AC1.W2.RS.
The diameter of the immobilized bioactive membrane is 2 mm and the mean applied activity is 1 unit/mm2.
AC*. W*. R* (*)
AC – amperometric sensor or electrode on corundum ceramic base
AC*– Sensor group reference number: 1-13
CC*. W* (*)
CC – conductometric electrode on corundum ceramic base
CC*– Sensor group reference number: 1-3
W* – Working electrode material
S – Alloy of Gold and Platinum
1 – Pure gold
2 – Pure platinum
3 – Pure silver
4 – Carbon(Graphite)
5 – Manually Microdispensed Carbon(Graphite) with Au+Pt alloy auxiliary electrode
R* – Reference electrode material
S – Silver
1 – Silver/Silver Chloride
2 – Silver covered by AgCl
(*) – Additional Technical specification
H – Heating of the sensor
T – Temperature sensing element