The Epsilon EClipseTM is the latest electrochemical analyzer manufactured by BASi. It is a potentiostat/galvanostat with a second working electrode for bipotentiostat measurements. Software with the most popular electrochemistry techniques is included with the instrument.
This System Includes:
- Basic Software – techniques include
- Cyclic Voltammetry/Linear Sweep Voltammetry
- Chronoamperometry/Chronocoulometry
- Controlled Potential Electrolysis
- DC Potential Amperometry
- Open Circuit Potential vs. Time
- Chronopotentiometry
- Basic Plus Software – techniques include
- Sampled Current Polarography
- Normal Pulse Voltammetry/Polarography
- Differential Pulse Voltammetry/Polarography
- Square Wave Voltammetry
- Stripping Voltammetry Techniques
- Double Potential Step Chronopotentiometry
- Methods Software – techniques include
- Prepare sequences with an unlimited number of techniques
- Insert trigger and delays in sequence
- Repeat sequence up to 9999 times
- Bi-Potentiostat – techniques include
- Multi-Channel Amperometry
- Multi-Channel Cyclic Voltammetry
- Multi-Channel Chronoamperometry
10’ USB Cable
Bi-Potentiostat cell lead cable
Potentiostat
- Channel 1 applied potential: Dynamic DAC: 16 bit, ±3.275 V @ 0.1 mV resolution or ±10 V @ 0.33 mV resolution
- Offset DAC: 8 bit, ±2.55 V @ 10 mV resolution
- Channel 2 applied potential: Dynamic DAC: 8 bit, ±2.55 V @ 10 mV resolution
- Compliance voltage: ±12 V
- Maximum current: 100 mA
- Bandwidth: > 100 kHz
- Bias current: < 1pA
- Electrometer input impedance: > 1012W
- (values for parameters other than applied potential are for both channels)
Current-to-voltage converter
- Full scale sensitivity: 1 nA*, 10 nA*, 100 nA, 1 µA, 10 µA, 100 µA, 1 mA, 10 mA,100 mA, (*secondary gain used)
- Secondary gain: x1, x10, and x100
- ADC resolution: 16 bit
- Sampling rate: 50 kHz (20 µs/conversion)
- Data length: up to 64k points – fast
Galvanostat
- Applied current: 50 pA – 50 mA
- Range: ±10 V ± 1 V
- Measured potential resolution: 0.02 mV
- Max leakage current: 30 pA
I/O
- Eout (applied potential)
- Iout (current to voltage output)
- Ein (input connection for external applied wave form)
- Cell Stand port (control of C-3 Cell Stand and CGME)
- Ordering Port (control of RDE-2)
- PC (USB) port
- Trigger out (TTL)
Minimum PC requirements
- Win 7 or higher
- Pentium III
- 512 MB RAM
- 50 MB hard drive space available
- USB port
Power requirements
- 120 or 240 VAC, 50/60 Hz, 60 VA
Dimensions & weight
- 15.75″ (40 cm) x 5.25″ (13 cm) x 12.75″ (32.5 cm)
- 17.5 lb (7.4 kg) (1 channel)