Çalışma Saatleri : 09:00 - 18:00

Epsilon EClipse

The Epsilon EClipseTM is the latest electrochemical analyzer manufactured by BASi. It is a potentiostat/galvanostat with a second working electrode for bipotentiostat measurements. Software with the most popular electrochemistry techniques is included with the instrument.

This System Includes:

  • Basic Software – techniques include
    • Cyclic Voltammetry/Linear Sweep Voltammetry
    • Chronoamperometry/Chronocoulometry
    • Controlled Potential Electrolysis
    • DC Potential Amperometry
    • Open Circuit Potential vs. Time
    • Chronopotentiometry
  • Basic Plus Software – techniques include
    • Sampled Current Polarography
    • Normal Pulse Voltammetry/Polarography
    • Differential Pulse Voltammetry/Polarography
    • Square Wave Voltammetry
    • Stripping Voltammetry Techniques
    • Double Potential Step Chronopotentiometry
  • Methods Software – techniques include
    • Prepare sequences with an unlimited number of techniques
    • Insert trigger and delays in sequence
    • Repeat sequence up to 9999 times
  • Bi-Potentiostat – techniques include
    • Multi-Channel Amperometry
    • Multi-Channel Cyclic Voltammetry
    • Multi-Channel Chronoamperometry

10’ USB Cable
Bi-Potentiostat cell lead cable

Potentiostat

  • Channel 1 applied potential: Dynamic DAC: 16 bit, ±3.275 V @ 0.1 mV resolution or ±10 V @ 0.33 mV resolution
  • Offset DAC: 8 bit, ±2.55 V @ 10 mV resolution
  • Channel 2 applied potential: Dynamic DAC: 8 bit, ±2.55 V @ 10 mV resolution
  • Compliance voltage: ±12 V
  • Maximum current: 100 mA
  • Bandwidth: > 100 kHz
  • Bias current: < 1pA
  • Electrometer input impedance: > 1012W
  • (values for parameters other than applied potential are for both channels)

Current-to-voltage converter

  • Full scale sensitivity: 1 nA*, 10 nA*, 100 nA, 1 µA, 10 µA, 100 µA, 1 mA, 10 mA,100 mA, (*secondary gain used)
  • Secondary gain: x1, x10, and x100
  • ADC resolution: 16 bit
  • Sampling rate: 50 kHz (20 µs/conversion)
  • Data length: up to 64k points – fast

Galvanostat

  • Applied current: 50 pA – 50 mA
  • Range: ±10 V ± 1 V
  • Measured potential resolution: 0.02 mV
  • Max leakage current: 30 pA

I/O

  • Eout (applied potential)
  • Iout (current to voltage output)
  • Ein (input connection for external applied wave form)
  • Cell Stand port (control of C-3 Cell Stand and CGME)
  • Ordering Port (control of RDE-2)
  • PC (USB) port
  • Trigger out (TTL)

Minimum PC requirements

  • Win 7 or higher
  • Pentium III
  • 512 MB RAM
  • 50 MB hard drive space available
  • USB port

Power requirements

  • 120 or 240 VAC, 50/60 Hz, 60 VA

Dimensions & weight

  • 15.75″ (40 cm) x 5.25″ (13 cm) x 12.75″ (32.5 cm)
  • 17.5 lb (7.4 kg) (1 channel)