Çalışma Saatleri : 09:00 - 18:00

TCM-K

Temperature measurement module

The TCM-K K-type thermocouple module is used to connect K-type thermocouples to the peripheral port of an Ivium potentiostat. In this way, temperatures can be measured and recorded simultaneously with the primary signal.

  • Suitable for K-type thermocouples
  • The input of the thermocouple module is not isolated from the potentiostat’s ground. If isolation of the thermocouple is required, a thermocouple with isolated tip should be used
  • Powered from the 5V supply available in the  peripheral port of the potentiostat
  • Temperature range: 10 to 900C
  • Measurement accuracy:
    • 10 to 400C is ±2C
    • >400C is ±1.6%
  • The TCM-K can be used in ambient temperatures: 0 to 50C
  • Output signal: analog output voltage of 5mV/C
  • Port interfacing: the voltage signal transfers to analog input1 of the potentiostat peripheral port